Technical Specifications
1.Bolus: 600ml/h
2.Flow rate range: 1-600ml/h
3.Volume limit: 1-9999 ml
4.Accuracy:±5
5.KVO flow rate:1ml/h, keep vein open(KVO) rate
6.Power supply: AC100-240V, 50/60 Hz, 25VA
7.Water Proof: IP ×I
8.Battery: Rechargeable lithium polymer battery,7.4 V,1650mAh
9.Electrical safety: compliance with the requirements of IEC 60601-1.
10.Max. Power consummation: 25W
11.Battery recharge: When the pump is connected to the AC power, the battery will automatically recharge (About 8
-14 hours to recharge fully.) Can run for more than 5 hours continuously after fully recharged.
12.Fuse Type:220V 2A×2,12V 2A×2
13.Display or information: Flow rate, volume limit, accumulated volume, power indicator light, bed No., air, occlusion, empty.
14.Alarm function: Infusion completion, occlusion, air bubble, low battery, control abnormal, no AC power supply, installation error.
15.Max. size of outer shell:120×140×190 mmlength× width× height
16.Max. weight:2.5 kg
17.Classification: Type B.
18.Outer shell material: ABS plastic
19.Operating condition:Environment temperature:+5- +40,atmosphere pressure 50
106kPa,related humidity 30%-90%
20.Storage and transport condition: Environment temperature -15-
+50,atmosphere pressure 50-106kPa,relative humidity 30%-90%.
21.Applicable infusion pipe: All standard infusion pipe(use "double dove" to test )
22.EMC: Complies with IEC/EN60601-1-2 and IEC/EN60601-2-24.